Pacemaker FAQ

From Computing and Software Wiki

(Difference between revisions)
Jump to: navigation, search
Lawford (Talk)
(Started entering FAQ contents.)
Newer edit →

Revision as of 20:46, 20 May 2008

This page contains answers to the Frequently Asked Questions about the Pacemaker.

Requirements

  1. Section 2.5.2 – Implant Phase
    7. Testing the system sensing and pacing efficacy.
    • What kind of test is made?
    • Which of these parameters are evaluated?
      A: See #6 in the section above 2.5.2
  2. Part 5.2 said URI is the min time between a ventricular event and the next ventricular pace. Then why is URI effective in modes without ventricular pacing nor sensing, such as AAT, AOO (Table 6, page 28)?
    A: “URI” should be “URL interval” in the question. For AAT, trigger paces must not be faster than URL. AOO is constant rate pacing at LRL. So long as LRL < URL (one of many implicit parameter constraints) AOO will pace slower than URL.
    The requirement could have been broadened to atrial-only modes. Alternatively, strict reading that URL does not apply to atrial-only modes is acceptable as well.

Hardware Reference Platform

Personal tools