Pacemaker FAQ
From Computing and Software Wiki
(Difference between revisions)
												
			
		|  (Started entering FAQ contents.) |  (Added more parts of the FAQ.) | ||
| Line 7: | Line 7: | ||
| #* Which of these parameters are evaluated?<br>'''A:'''	See #6 in the section above 2.5.2 | #* Which of these parameters are evaluated?<br>'''A:'''	See #6 in the section above 2.5.2 | ||
| # Part 5.2 said URI is the min time between a ventricular event and the next ventricular pace. Then why is URI effective in modes without ventricular pacing nor sensing, such as AAT, AOO (Table 6, page 28)?<br>'''A:''' “URI” should be “URL interval” in the question.  For AAT, trigger paces must not be faster than URL.  AOO is constant rate pacing at LRL.  So long as LRL < URL (one of many implicit parameter constraints)  AOO will pace slower than URL.<br> The requirement could have been broadened to atrial-only modes.  Alternatively, strict reading that URL does not apply to atrial-only modes is acceptable as well. | # Part 5.2 said URI is the min time between a ventricular event and the next ventricular pace. Then why is URI effective in modes without ventricular pacing nor sensing, such as AAT, AOO (Table 6, page 28)?<br>'''A:''' “URI” should be “URL interval” in the question.  For AAT, trigger paces must not be faster than URL.  AOO is constant rate pacing at LRL.  So long as LRL < URL (one of many implicit parameter constraints)  AOO will pace slower than URL.<br> The requirement could have been broadened to atrial-only modes.  Alternatively, strict reading that URL does not apply to atrial-only modes is acceptable as well. | ||
| + | # 2.5.3 - Predischarge Follow-up (Page 12)<br>2. Reprogramming to final pre-discharge value | ||
| + | #* Is there any special values for parameters?<br>'''A.'''	No. This just means that the doctor can change parameters after implant, but before the patient leaves the hospital. | ||
| + | # 3.6.1 - Permanent State (Page 17)<br>The normal pacing parameters programmed shall be used in the permanent brady state.  | ||
| + | #* Which are these normal pacing parameters?<br>'''A.'''	The ones that are currently programmed. | ||
| + | # 3.6.2 - Temporary Bradycardia Pacing<br>The temporary brady parameters programmed shall be used in the temporary brady state.  | ||
| + | #* Which are these temporary brady parameters?<br>'''A.'''	An identical set of parameters to the permanent parameters, but changes in temporary parameters do not affect permanent parameters.<br> | ||
| + | The temporary state shall be capable of being used to temporarily test various system parameters or provide patient diagnostic testing. | ||
| + | #* Which parameters are tested? <br>'''A.'''	Any or all parameters that the physician wishes to try temporarily. | ||
| + | #* What is the "patient diagnostic testing".<br>'''A.'''	Testing impedance/amplitude/threshold. | ||
| + | |||
| + | |||
| + | |||
| == Hardware Reference Platform == | == Hardware Reference Platform == | ||
Revision as of 02:23, 21 May 2008
This page contains answers to the Frequently Asked Questions about the Pacemaker.
Requirements
-  Section 2.5.2 – Implant Phase
 7. Testing the system sensing and pacing efficacy.- What kind of test is made?
-  Which of these parameters are evaluated?
 A: See #6 in the section above 2.5.2
 
-  Part 5.2 said URI is the min time between a ventricular event and the next ventricular pace. Then why is URI effective in modes without ventricular pacing nor sensing, such as AAT, AOO (Table 6, page 28)?
 A: “URI” should be “URL interval” in the question. For AAT, trigger paces must not be faster than URL. AOO is constant rate pacing at LRL. So long as LRL < URL (one of many implicit parameter constraints) AOO will pace slower than URL.
 The requirement could have been broadened to atrial-only modes. Alternatively, strict reading that URL does not apply to atrial-only modes is acceptable as well.
-  2.5.3 - Predischarge Follow-up (Page 12)
 2. Reprogramming to final pre-discharge value-  Is there any special values for parameters?
 A. No. This just means that the doctor can change parameters after implant, but before the patient leaves the hospital.
 
-  Is there any special values for parameters?
-  3.6.1 - Permanent State (Page 17)
 The normal pacing parameters programmed shall be used in the permanent brady state.-  Which are these normal pacing parameters?
 A. The ones that are currently programmed.
 
-  Which are these normal pacing parameters?
-  3.6.2 - Temporary Bradycardia Pacing
 The temporary brady parameters programmed shall be used in the temporary brady state.-  Which are these temporary brady parameters?
 A. An identical set of parameters to the permanent parameters, but changes in temporary parameters do not affect permanent parameters.
 
 
-  Which are these temporary brady parameters?
The temporary state shall be capable of being used to temporarily test various system parameters or provide patient diagnostic testing.
-  Which parameters are tested? 
 A. Any or all parameters that the physician wishes to try temporarily.
-  What is the "patient diagnostic testing".
 A. Testing impedance/amplitude/threshold.
 
-  Which parameters are tested? 